451B

Operators Manual

20

Troubleshooting

Caution

For safe operation and maintenance of the Product, use caution

when working on this Product. If the device indicates a

relatively high radiation rate, there is a possibility that the

Product has been contaminated. High rates can quickly cause

the operator’s cumulative exposure to increase with the

potential for injury.

Note

Install fresh batteries prior to performing any calibration on this
product.

Self-Test Failure

If the Survey Meter fails self-test, it remains locked on the firmware revision

number. See the How to Contact Fluke section.

High Background Readings

The desiccant must be renewed and replaced periodically. See the How to

Contact Fluke section.

Erratic Readings

Check for broken ion chamber window or loose collector. Check for loose

chamber bias connector.