Ion Chamber Survey Meter

Maintenance

17

Battery Replacement

 Warning

To prevent possible electrical shock, fire, or personal injury:

Batteries contain hazardous chemicals that can cause

burns or explode. If exposure to chemicals occurs, clean
with water and get medical aid.

Do not disassemble the battery.

Repair the Product before use if the battery leaks.

Be sure that the battery polarity is correct to prevent battery

leakage.

Do not short the battery terminals together.

Do not disassemble or crush battery cells and battery

packs.

Do not keep cells or batteries in a container where the

terminals can be shorted.

Do not put battery cells and battery packs near heat or fire.

Do not put in sunlight.

The LOW BAT message will appear on the display approximately 6 hours prior to

the Survey Meter becoming inoperable. To ensure that the Survey Meter

operates to specification, change the batteries within 1 hour after the LOW BAT

message appears. The Survey Meter will function on one battery for

approximately 100 hours, allowing replacement of one battery at a time if the

Survey Meter must remain operational during battery changeover. Use either

regular or alkaline batteries for replacement purposes.
Batteries are accessible from the rear of the Survey Meter.