Invasive blood pressure, Non-invasive blood pressure – Fluke Biomedical ProSim 4 Getting Started User Manual
Page 31

Vital Signs Simulator
Detailed Specifications
19
Invasive Blood Pressure
Channels ................................................................ 1 electrically isolated from all other signals
BP Output ............................................................... Circular DIN 5-Pin
Input/output Impedance ........................................ 300
Ω
±
10 %
Exciter Input Range ............................................... 2.0 to 16.0 V peak
Exciter-Input Frequency Range ........................... DC to 5000 Hz
Transducer Sensitivity .......................................... 5
μ
V/V/mmHg
Pressure Accuracy ................................................
±
(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation only
Static Pressure ...................................................... 0, 80, 160, and 250 mmHg
Dynamic Waveforms
Synchronization ................................................... To ECG heartrate
Chambers simulated systolic/diastolic pressure:
Type
IBP (Arterial)
IBP (Left Ventrical)
Adult 60/30
60/0
Adult 120/80
120/0
Adult 150/100
150/0
Adult 200/150
200/0
Neonatal 35/15
35/0
Neonatal 70/40
70/0
Non-Invasive Blood Pressure
Pressure Units ....................................................... mmHg
Manometer (Pressure Meter)
Range .................................................................. 10 to 400 mmHg
Resolution ........................................................... 0.1 mmHg (for display purposes)
Accuracy..............................................................
±
(1 % reading +1 mmHg)
Pressure Source .................................................... Inflation bulb or device under test