Vital Signs Simulator

Detailed Specifications

19

Invasive Blood Pressure

Channels ................................................................ 1 electrically isolated from all other signals
BP Output
............................................................... Circular DIN 5-Pin
Input/output Impedance ........................................ 300

Ω

±

10 %

Exciter Input Range ............................................... 2.0 to 16.0 V peak
Exciter-Input Frequency Range ........................... DC to 5000 Hz
Transducer Sensitivity .......................................... 5

μ

V/V/mmHg

Pressure Accuracy ................................................

±

(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation only

Static Pressure ...................................................... 0, 80, 160, and 250 mmHg
Dynamic Waveforms

Synchronization ................................................... To ECG heartrate

Chambers simulated systolic/diastolic pressure:

Type

IBP (Arterial)

IBP (Left Ventrical)

Adult 60/30

60/0

Adult 120/80

120/0

Adult 150/100

150/0

Adult 200/150

200/0

Neonatal 35/15

35/0

Neonatal 70/40

70/0

Non-Invasive Blood Pressure

Pressure Units ....................................................... mmHg
Manometer (Pressure Meter)

Range .................................................................. 10 to 400 mmHg
Resolution ........................................................... 0.1 mmHg (for display purposes)
Accuracy..............................................................

±

(1 % reading +1 mmHg)

Pressure Source .................................................... Inflation bulb or device under test