ProSim™ 4
Getting Started

18

Ventricular Fibrillation .......................................... Coarse or fine
Transvenous Pacer Pulse .................................... 75 BPM, left arterial, 3 mV amplitude on lead II, Accuracy

±

10 %, 1.0 ms width

2nd Degree AV Block ............................................ Type 1
3rd Degree AV Block
Asystole

ECG-Performance-Testing

Amplitude .............................................................. 1 mV. Other leads are proportional to Lead II (reference lead) in percentage per:

Lead I .............................................................. 70
Lead II ............................................................. 100
Lead III ............................................................ 30
Lead V1 ........................................................... 24
Lead V2 ........................................................... 48
Lead V3 ........................................................... 100
Lead V4 ........................................................... 120
Lead V5 ........................................................... 112
Lead V6 ........................................................... 80

Square Wave ......................................................... 60 ms at 2.0 Hz

Respiration

Rate ........................................................................ 0 (OFF), 10 to 100 BrPM in 10 BrPM steps
Impedance Variations (

Δ

Ω

) ................................. 1

Ω

Accuracy Delta ......................................................

±

(10 % + 0.05 ohm)

Baseline ................................................................. 500

Ω

to circuit common, giving 1000

Ω

between any two leads

Accuracy Baseline ................................................

±

5 %

Respiration Lead ................................................... LA or LL (default)