Fluke Biomedical 960SF-220 User Manual
Page 18

Victoreen 960SF-220,221,230,231
Operators Manual
A-2
J2 Pin Designation
* Pin
Description
* Pin
Description
A
Analog Out +
a
SIG- (2)
B GND
b SIG+
(1)
C Not
Used
c GND
D
Not Used
d
Not Used
E Not
Used
e SIG+(2)
F Not
Used
f GND
H Not
Used
h GND
J Not
Used
i GND
K
Not Used
k
Not Used
L GND
m Not
Used
M
Not Used
n
Not Used
N
Not Used
D
Not Used
P GND
r K3
(ALARM)
R
SIG- (3)
s
Relay Common
S
Counter Enable (Not
Used)
t K5
(HORN)
T GND
u K4
(CHECK
SOURCE)
U
SIG+ (4)
Y
K1 (FAIL)
V
SIG+ (3)
w
K6 (BEACON)
W GND
x K2
(WARN)
X SIG-
(1)
Y SIG-
(4)
Z GND
J3 Pin Designation
Pin Description
Pin Description
1 GND
14
FAIL/ACK
SW
2
GND
15
C/S SW N.O.
3
-
16
C/S SW N.C.
4 -
17
-
5 -
18
TEST/NORMAL
SW
(TP)
6 -
19
+15
VDC
7 Warn
Light
20
+15
VDC
8
C/S Light
21
SIG IN (TEST) (TP)
9
Scare Light
22
SHUT DOWN LED
10
Scare Light
23
+5 VDC
11
Alarm Light
24
+5 VDC
12
Scare Light
25
HV Test Point
13 Fail
Light
26 -
N.O. - Normally Open Contacts