4 functional description – Fluke Biomedical 943-25 User Manual
Page 6

Victoreen 943-25, 943-26
Operators Manual
1-2
Response Ratio:
1:1
943-25, 943-26, 943-25T, 943-26T
10:1
943-25A, 943-26A, 943-25TA, 943-26TA
100:1
943-25B, 943-26B, 943-25TB, 943-26TB
End Window Material
Mylar
943-25, 943-25A, 943-25B, 943-26, 943-26A, 943-26B
Titanium 943-25T, 943-25TA, 943-25TB, 943-26T, 943-26TA, 943-26TB
Preamplifier
Maximum Pressure
30 psi
(Face Exposure)
Rise Time
< 60 ns
Input Impedance
> 50 kilohms
Output Coupling
AC
Voltage Gain
6 V/V
Configuration
Voltage sensitive
Output Polarity
Negative
Output Impedance
50 ohms
Maximum Cable Length
1500 ft
Maximum Pulse Amplitude @ -6 V
Maximum Cable Length
Dead Time
Approximately 10 us @ 8 ft cable length
Maximum Count Rate
10
7
The beta detectors have been assembled with parts selected for the reliability required in a nuclear
application. Any unauthorized repairs made to the detectors utilized in nuclear applications will void the
safety-related rating. Safety-related detectors must be returned to Fluke Biomedical, Radiation
Management for authorized, qualified (ANSI 45.2.6, Skill Level II) service.
1.4 Functional Description
The Victoreen beta scintillation detector is shown in block diagram form in Figure 1-1. The detector
consists of a scintillation disc, a photomultiplier tube, a preamplifier assembly, and various interconnecting
cables connecting the detector, readout, and the detector power supply that is located at the ratemeter or
scaler module.
Figure 1-1. Block Diagram, 943 Series Beta Detector