Maternal ecg – Fluke Biomedical PS320 User Manual
Page 14

PS320
Users Manual
6
Table -1. Fetal ECG Patterns (cont.)
Pattern Description
EXAGGERATED DECELE. Exaggerated
deceleration
NON_UNIFORM DECELE.
Non-uniform deceleration (2 different shapes)
VAR. DECELERATION (U)
Variable deceleration “U” shaped
VAR. DECELER. TACH
Variable deceleration with high rate BPM
VAR. DECELER (V)
Variable deceleration “V” shaped
VAR. DECEL. (POST)
Variable deceleration post exaggerated
VAR. DECELERATION Variable
deceleration
DECELER. (POSITION) Variable
deceleration
with position changes
LONG DECELERATION Long
deceleration
COMPENSATORY ACCEL. Compensatory
acceleration
Maternal ECG
Maternal ECG static rates.....................................60, 80, 100, 120, 140, 160 BPM
Maternal ECG sensitivity (amplitude)..................0.5 mV, 1 mV, and 2 mV
Select pattern during Trend #1 selection
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)