Clean the product (outside), Clean the product (inside), Replaceable parts – Fluke Biomedical IDA-1S User Manual
Page 29

Infusion Device Analyzer
Product Maintenance
19
Clean the Product (Outside)
To clean the outside of the Product, disconnect from the power supply and use
only a damp cloth with mild detergent.
Clean the Product (Inside)
It is possible that microbial growth can become present in the measuring module
of the Product. It is recommended that you clean the fluid paths at 3 month
intervals. To clean the inside of the Product, inject 20 ml of a warm water and
detergent solution into the Fluid Inlet Port. After 5 minutes, flush with clean water.
Always pass water from the fluid inlets to the outlets.
Replaceable Parts
Table 11 lists the replaceable parts for the Product.
Table 11. Replaceable Parts
Item
Fluke Biomedical
Part Number
IDA-1S Infusion Device Analyzer
4468525
Hydrograph software and IDA-1S Users Manual on CD
4418071
Plastic syringe (20 ml)
4497350
3-way Luer plastic stop-cock
4480194
Drain tube (1 m)
4478942
Micro-90
(225 ml)
4541948
USB Cable
1740487
Power supply, 4-plug mains adapter kit for US, UK, EURO, AUS
2461300
IDA-1S power supply cord without mains adapters, universal wall mount
4329971
NiMH Battery
4481150
IDA-1S Getting Started Manual
4426198