Safety Information
A Warning identifies conditions and procedures that are dangerous to the user.
A Caution identifies conditions and procedures that can cause damage to the
Product or the equipment under test.
Warning
To prevent possible electrical shock, fire, or personal injury:
• Read all safety information before you use the Product.
• Use the Product only as specified, or the protection
supplied by the Product can be compromised.
• Remove the batteries if the Product is not used for an
extended period of time, or if stored in temperatures above
50
°C. If the batteries are not removed, battery leakage can
damage the Product.
• The battery door must be closed and locked before you
operate the Product.
• Replace the batteries when the low battery indicator shows
to prevent incorrect measurements.
• Carefully read all instructions.
• Do not touch voltages >30 V ac rms, 42 V ac peak, or
60 V dc.
• Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
• Examine the case before you use the Product. Look for
cracks or missing plastic. Carefully look at the insulation
around the terminals.
• Use this Product indoors only.
• Use only the mains power cord and connector approved for
the voltage and plug configuration in your country and rated
for the Product.
• Replace the mains power cord if the insulation is damaged
or if the insulation shows signs of wear.
• Use only the external mains power supply included with the
Product.
• Remove all probes, test leads, and accessories before the
battery door is opened.
• Disable the Product if it is damaged. Remove the batteries to
disable the Product.
• Do not use the Product if it is damaged.
• Do not use the Product on infusion devices that are
attached to patients.