Introduction, 1 product description, 2 specifications – Fluke Biomedical 07-479 User Manual
Page 5

Introduction
Product Description
1
1-1
Section 1
Introduction
1.1 Product Description
The Model 07-479 Dental kVp Meter is designed to measure the effective peak potential applied to dental
type x-ray tubes. The measurement is made without direct connection to the x-ray unit being tested. The
kVp Meter is a battery powered, self-contained instrument designed for portable use.
The kVp meter uses two differentially filtered detectors whose ratio of integrated outputs is calibrated over
the range of 45 to 90 kVp. Measurement results are presented on an easy to read 3 ½ digit liquid Crystal
Display. Status is indicated by decimal point positioning on the readout. Low battery condition is indicated
by LO BAT displayed on the readout.
1.2 Specifications
Range
45 to 90 kVp
Accuracy
± 3 kV or 3%, whichever is greater.
Calibrated with NBS traceable voltage divider.
Reproducibility
± 0.5 kV
Resolution 0.1
kV
Distance Dependence
Negligible for SSD of 20 cm or greater
Angle Dependence
Negligible for 10 degrees or less
Position Dependence
Negligible for misalignment of 5 cm or less at 30 cm
Calibration Period One
year
Minimum Requirements
5 mAs @ 45 kVp (20 cm)
0.03 mAs @ 90 kVp (20 cm)
Power Requirements
9 V Alkaline Battery, MN1604A or equivalent
Display
3 ½ Digit LCD
Low Battery Indication: LO BAT
High kV Indication: Left blinking decimal
Lo kV Indication: Middle blinking decimal
High Intensity Indication: Right blinking decimal and non-zero display
Low mAs Indication: Right blinking decimal and zero display
LED Indicator
Auto Reset, new exposure detected, and last reading cleared.
Controls
Front Panel: Power switch and phase selection switch.
Output Connector
BNC Connector on rear panel for waveform output.