Introduction, 1 general description, 2 application – Fluke Biomedical 6000-530B User Manual
Page 5: 3 specifications

Introduction
General Description
1
1-1
Section 1
Introduction
1.1 General Description
The Model 6000-530B Image Intensifier Ion Chamber is a low profile ion chamber designed to measure
diagnostic
x-rays.
1.2 Application
The Model 6000-530B Image Intensifier Ion Chamber is specifically designed for measurement of
exposure rate at the input phosphor of fluoroscopic image intensifier tubes. Its sensitive area and form
factors have been adapted for placement inside the spot film tray of common image intensifier systems.
Also included is a detachable handle for easy insertion and removal. When the host instrument is a Model
6000 series or a Model 4000M+, precise measurements may be made at low exposure rates (down to 10
μR/min) if the Model 6000-530B Preamplifier is used (See the Model 06-524- 2000 manual for operating
instructions).
1.3 Specifications
Chamber
Volume
150 cm
3
Nominal Sensitivity
45 nC/R
Housing Material
Acrylic
Housing Color
Clear
Housing Size
6.5 in diameter (16.5 cm diameter)
Surface Area
110 cm
2
, 5.5 in diameter
Weight
30 oz (840 g)
Handle Length
16.5 in (41.9 cm)
Height
0.720 in
Cable Length
10 ft (3 m)
Cable Termination
BNC/Banana
Window Thickness
.110 in (2.79 mm)
Total Attenuation
0.275 mm AI equivalent @ 30 keV
Operating Voltage
200 – 300 V DC
Compatible With
Model 6000 Series, 4000M+, & RAD-CHECK® PLUS