Flowserve 1878 Double-Disc Gate Valve User Manual
Page 14

14
9.0
MAINTENANCE INSTRUCTIONS (Continued)
 
 9.4 
Trouble Shooting
 
 
A.
EXCESSIVE HANDWHEEL (136) EFFORT OR BINDING
 
 
Excessive handwheel effort or binding would indicate that the Stem (025) 
either needs to be lubricated; or the Gland Flange (133) is too tight as a 
result of the Gland Stud Nuts (234) being tightened unevenly. Lubrication 
(Para. 7.3) should be checked at regular intervals of 6-months. To tighten 
the Gland Stud Nuts evenly, they should be loosened and retightened in an 
alternating fashion. 
 
 
B.
LEAKAGE BETWEEN THE DISC (004) AND SEAT (013)
 
 
This could be an indication that there is foreign matter on the seating 
surfaces in which case the valve should be opened and closed ONE TIME in 
an attempt to dislodge any matter that may have inadvertently lodged there. 
DO NOT OPEN AND CLOSE THE VALVE MORE THAN ONCE AS THIS 
MAY CAUSE PERMANENT DAMAGE TO THE SEATING SURFACES IF 
THE FOREIGN MATTER IS SECURELY LODGED IN PLACE. Instead, 
disassemble the valve and remove the sources of the trouble. If no foreign 
matter is found, inspect the seating surfaces of the valve for signs of a 
scarred or damaged seat - in which case the seating surfaces of the Disc 
(004) and Seat Ring (013) should be lapped until no visible defects remain. 
(Refer to Para. 9.3) 
 
 
C.
LEAKAGE BETWEEN THE BODY (001) AND BONNET (002)
 
 
This would be a good indication that the Gasket (100) is worn out or 
damaged. Replacement of gaskets is set forth in Para. 9.1. Another source 
of the trouble might be that the Bonnet Studs and Nuts are loose - if this 
should be the case - tighten them securely. 
 
 
D.
LEAKAGE AROUND STEM (025) AND THRU THE STUFFING BOX
 
 
Worn Packing (110, 112). Replace the packing following the instructions set 
forth in Para. 9.2. 
