beautypg.com

Flowserve X-200 User Manual

Page 5

background image

2.2 Install the pump back plate (stuffing box), Figure 4. Position the

X-200 seal gland flush connection, port a in Figure 7, in the

3:00 o’clock position for normal installations with CW rotation

(9:00 o'clock for CCW rotation) for optimum heat removal. Alternate

positioning of the gland may be necessary with some pump

designs to permit seal piping to clear the bearing housing. Refer

to the assembly drawing that is supplied with the seal for proper

positioning. Bolt the back plate in place on the bearing housing.

2.3 Position the X-200 cartridge gland with the gland gasket in

place against the seal chamber (stuffing box) face, install flat

washers on the studs to prevent misalignment, and tighten the

gland nuts evenly, cross staggering the adjustment of the nuts.

Do not over tighten.

2.4 Assemble the pump, adjust the bearings, set the impeller and

connect the coupling so that the shaft is in its operating axial

position. Any subsequent axial adjustment of the shaft requires

resetting of the seal. Connect pump piping, allow no pipe strain

on the pump casing.

5

2.1 Lubricate the shaft or sleeve OD lightly with silicone lubricant

provided and slide the complete X-200 cartridge seal onto

the shaft, Figure 3, with the end with the setting devices toward

the bearing housing.

Install pump backplate

Figure 4