Specifications – Electro-Chemical Devices (ECD) L20 C User Manual
Page 11
Specifications
Display
Range
Resolution
Accuracy
Conductivity
K=0.01
0.000 to 1.999uS/cm
2.00 to 19.99uS/cm
0.001uS/cm
0.01uS/cm
±
0.5% FS
Conductivity
K=0.01
0.00 to 19.99uS/cm
2.0 to 199.9uS/cm
0.01uS/cm
0.1uS/cm
±
0.5% FS
Conductivity
K=0.01
0.0 to 199.9uS/cm
200 to 1999uS/cm
2.00 to 19.99mS/cm
0.1uS/cm
1uS/cm
0.01mS/cm
±
0.5% FS
Conductivity
K=0.01
0 to 1999uS/cm
2.00 to 19.99mS/cm
20.0 to 199.9mS/cm
1uS/cm
0.01mS/cm
0.1mS/cm
±
0.5% FS
Temperature 0.0 to 100.0 °C
0.1 °C
±
0.2°C
Reference Temperature
15.0 to 25.0 °C
Temperature Coefficient
0.0% to 4.0%
Cell Constant
0.01; 0.10; 1.00; 10.0
TDS Constant Range
0.30 to 1.00, default at 0.65
Power
Six “AAA” Batteries
Calibration Back-up
EEPROM
Audio Feedback
All Touch Keys
Display
22mm : 14.5mm high LCD
Ambient Temperature
0 to 50 °C
Relative Humidity
up to 90%
Case
IP54
Dimensions
150mm x 203mm x 72mm
Weight
504 grams(Batteries included)
11