4. Rinse the CELL at least 3 times with the water sample you will be testing -

rinsing minimizes the potential for cross contamination from a previous test.

Finally, fill cell to capacity (4mL) with the water sample.

5. Press the ZERO/ON button. The cursor will move across the display followed

by 0.00 PPM. Sample is ready for testing.

6. Dip the HCA1 Micro Strip PA (DPD-4) into the CELL and immediately press

READ. This starts the 20 SECOND countdown timer. During this time move the

strip in a gentle back and forth motion (approx. 2 strokes/Sec). Remove and

discard the strip after “1” on the display disappears. When testing outdoors

(in sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing

and Reading the sample.

7. The cursor will move across the display while the meter prepares to measure

the sample. Record result displayed (this result is automatically stored in PA5

Menu).

After testing is completed, rinse cell immediately.

MENU HR8 High Range Chlorine Test Procedure

1. Remove (1) Micro HR Strip from the bottle before beginning the test. Place it

in a dry convenient place and recap the test strip bottle.

2. Press the ZERO/ON button to power the meter on; the display will show all

annunciators, then the current MENU selection, followed by the last reading.

3. Press and re-press the MENU button until the display shows the parameter

HR8.

4. Rinse the CELL at least 3 times with the water sample you will be testing -

rinsing minimizes the potential for cross contamination from a previous test.

Finally, fill cell to capacity (4mL) with the water sample.

5. Press the ZERO/ON button. The cursor will move across the display followed

by 0.00 PPM. Sample is ready for testing.

6. Dip the HCA1 Micro Strip HR into the CELL and immediately press READ. This

starts the 20 SECOND countdown timer. During this time move the strip in a

gentle back and forth motion (approx. 2 strokes/Sec). Remove and discard the

strip after “1” on the display disappears. When testing outdoors (in sunlight),

for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading

the sample.

7. The meter will begin counting up for 120 seconds, at the end of which the

cursor will move across the display while the meter prepares to measure the

sample. Record result displayed (this result is automatically stored in HR8

Menu).

After testing is completed, rinse cell immediately.

10