Digi-Star Stockweigh 300 User Manual
Page 22

Troubleshooting
18
SW300 User’s Manual
D3734-US REV E
Indicator Does Not Respond When You Step on the Scale
The other load cell is defective.
Check its cable for cuts or flattened
areas or send it back for repair.
Check the defective load cell by
connecting it to the indicator and
verify that it does not weigh.
YES
NO
The other load cell is defective. Check
its cable for cuts or flattened areas and
send it back for repair.
Both load cells are defective. The
indicator is probably OK. Check the
cables for damage and send both
load cells in for repair.
The indicator is defective. Send it in for
repair. Note: Load cells may also be
defective. Check the load cells by
attaching to a good working indicator.
YES
YES
NO
NO
Disconnect one of the load cells and stand over the connected load cell.
Does the indicator display a reading about twice your weight?
Disconnect the first load cell and connect
the other load cell to the same connector
on the indicator. Does the indicator display
a reading about twice your weight?
Perform the paper clip test. Take two paper clips and carefully insert one
into pin 2 and the other in pin 3 of either load cell connector jack on the
indicator. Hold one paper clip in each hand (if the touch it will not damage
the indicator). Does the reading on the indicator jump around?