Get your calibration number – Digi-Star EZ 4600 Manual USB User Manual
Page 43

Re-Calibrating Your Scale
D3843-US REV B
EZ4600 User’s Manual
39
SETTING
[display]
D.A.N
NO.
OPTIONS [displayed]
BOLD=DEFAULT
DESCRIPTION
CALIBRATION
SETUP NUMBER
(SETUP)
871
Quick entry method
selects weigh method 1-
4lbs, 5-8 kg, gain 1-9,
display counts 1-9 and
capacity *1000
CALIBRATION
NUMBER
(CAL)
872
Weight displayed at
0.4mV/V for these load
cells
RE-CALIBRATING YOUR SCALE
To re-calibrate your scale and make it even more accurate, document at
least 3 to 6 loads of varying sizes and measure the actual weight of all loads
on a certified scale.
Reading Too High
If the Indicator is reading higher than the certified scale, then the calibration
number is high and should be decreased proportionally.
Reading Too Low
If the Indicator is reading lower than the certified scale, then the calibration
number is low and should be increased proportionally.
Get your Calibration
Number
1. Enter
872
and press . The
calibration (
CAL
) number will
display. Example
CAL = 24280.
1
2 4 2 8 0
Press help for
more information
1
TOTAL CERTIFIED WEIGHT
TOTAL INDICATOR WEIGHT
X CURRENT CAL NUMBER=NEW CAL NUMBER