Short Form Calibration
14
GT400 User’s Manual
D3831 Rev B
SHORT FORM CALIBRATION
The Short Form Setup & Calibration procedure allows you to change “SETUP”
and “CAL” numbers of indicator.
Do not attempt to calibrate scale if indicator is not reading stable weights.
Calibration procedure will not fix instability, inconsistencies, or flashing
"RANGE" messages.
Obtain Current Set-up and Calibration Number
Write down current SETUP and CAL
numbers of your GT 400 indicator.
These numbers are displayed during
Self Test.
SETUP NUMBER
To run self test with indicator ON:
1. Press until LB symbol starts
to blink. Release, then press
to start Self Test.
2. Press to “pause” the Self-Test
while numbers are displayed.
3. Press again to allow self-test
to complete normally.
SETUP # _______________
CAL # ________________
Following is a list of functions that are controlled by the “SETUP” number:
Weigh Method (
W MTHD
) Gain
Display Units (
LB-KG
)
Scale
Capacity
Display Counts (
COUNT
)
1
3
2
1