beautypg.com

2 phase changes of fluid, 3 printing your results – Bronkhorst CoriCalc User Manual

Page 11

background image

9.17.056

page 11

4.1.3.4

Maximum pressure outlet (sensor only)

When this option has been selected, the maximum pressure outlet available after the sensor at a given flow and inlet
pressure can be calculated.

4.2

Phase changes of fluid

If FLUIDAT® has been installed on the same PC, CoriCalc will use its fluid properties to check the phase of the selected
fluid for a certain pressure and temperature. Normally it is not recommended to have phase changes in either the
sensor nor the valve. Besides the cooling effect, also the working of the sensor and valve will be far from optimal.
Therefore, phase transitions in the instruments should be avoided. A transition from liquid phase into gas phase will
have a strong cooling effect on both valve and sensor and will be indicated with the blue pictogram as shown below.













4.3

Printing your results

Under the menu item File >> Print, it is possible to send a screen shot of the results directly to a printer.
The comment line at the bottom of the screen enables the user to add a comment about e.g. the application or the
customer.