4.5 T

EST COMMUNICATION

For maintenance and service operations, it is often required to read from or write to a certain parameter. FlowDDE

contains two test facilities: one to test FLOW-BUS and DDE communication and one to test lower level ProPar

communication.

4.5.1 Test FLOW-BUS and DDE

Open the FlowDDE test form from the menu FLOW-BUS >> Test FLOW-BUS and DDE or F6. The following window

appears.

The window consists of two sections. At the Test FLOW-BUS section parameters can be read and written directly to

the instrument, regardless of the DDE behaviour that a separate poll command is required to read a parameter value

and only changes are written to the instrument. The DDE value is updated on read, but not on write.

On the contrary, at the Test DDE section parameters can be read and written like any DDE client would do. Pressing

Read shows the current value at the DDE level, but the actual instrument value is not read and may be different. At

Write the value is written to the DDE level and, because of the nature of DDE, only if it is different than the previous

value, the value is written to the instrument.

At the Test DDE section links

to client applications, e.g.

Microsoft Excel, can be easily

made. Select the channel

and parameter to which a

link is to be made.

Press Copy link to copy the

link to the clipboard. In the

client application, use the

Paste Special option and

select Paste link to paste the

link (e.g. Paste Special in

Microsoft Excel, see image).

Any parameter value change

will be sent to the client

application.

The other way around is

possible as well. Copy a link

from any application (e.g.

copy a cell from Excel). In the test form, press Paste. Any change in the application’s field result in the parameter

value in FlowDDE to be updated. Note that when the communication is closed, all links will be terminated and will not

be restored when the communication is opened again.