2c. Environmental ratings
EL-FLOW Prestige instruments are suitable for use in conditions between -10...70 °C and relative humidity of
10...90 %RH, unless specified otherwise. Please note that EL-FLOW Prestige instruments have an ingress protection
rating of IP-40, implying that the electronics housing and electrical connection do not offer any protection against
moist environments.
3. Check if system piping is clean
For reliable measurement always make sure the fluid stream is clean.
Use filters to assure a moisture-, oil- and particle-free gas stream.
Recommended pore size: 5 µm. If back flow can occur, a downstream
filter and a check valve are recommended too. For high flow rates select
a suitable filter size, to avoid too high pressure drop.
4. Mount/install instrument properly
Install the EL-FLOW Prestige Meter/Controller in the line, in accordance
with the direction of the FLOW arrow. The arrow for flow direction is
indicated on the body of the instrument. Tighten the fittings according
to the instructions of the supplier of the fittings. The use of Swagelok
RS-type stainless steel adapters is recommended.
Avoid installation in close proximity of mechanic vibration and/or heat sources.
5. Leak check
Check the system for leaks before applying (fluid) pressure. Especially if toxic, explosive or other dangerous fluids
are used!
6. Electrical connection
Electrical connections must be made with a standard cable or according to
the EL-FLOW Prestige hook-up diagram. The factory installed 9-pin sub-D
settings are indicated on the instrument label. Make sure that the
power supply is suitable for the indicated power ratings, indicated on the
instrument label, and that double or reinforced insulation is used for the
power supply. For a more detailed hook-up diagram see the different
variations under 'SCOPE OF THIS GUIDE'.
7a. Analog / Local operation
Connect the EL-FLOW Prestige to the power supply/readout unit using a
cable with 9-pin sub-D connector.
7b. Digital RS232 operation
Digital operation over RS232 can be established when using
the following setup. Connecting the instrument with a RS232
cable or an RS232 cable with a USB to RS232 converter to a PC
will allow you to use (free) Bronkhorst® software for Windows,
such as FlowDDE and FlowPlot. Make sure that the instrument
back-side label indicates RS232 settings and apply the proper
Baud rate settings. If the instrument is not set for RS232
communication, please refer to the Instruction Manual EL-
FLOW Prestige series (9.17.084) for switching to RS232
communication settings.