beautypg.com

Introduction 3 preparing for use, 0 introduction, 0 preparing for use – Bluelab Combo Meter User Manual

Page 3: English

background image

English

3

1.0 Introduction

The Bluelab Combo Meter has three press buttons; ‘pH / calibrate’, ‘nutrient / units’

and ‘temp / units’. The buttons have a short press and long press function. A short

press means a button is released in about one second. The long press is a button

being held for at least three seconds and released when the display starts fl ashing.

Turning the Combo Meter on

A short press of any button turns the Combo Meter on. The Combo Meter

automatically turns off after four minutes if no buttons are pressed. If the Combo

Meter turns off before the reading is taken, a short press of any button will turn the

Combo Meter on again.

1

2.0 Preparing for use

The following tasks must be performed before the Bluelab Combo Meter is used for

the fi rst time.

Insert batteries.
See section 6.0.

Connect pH probe
Connect the pH probe to the pH meter by lining

up the lugs of the BNC fi ttings. Fasten securely

by pushing the pH probe connector on and

twisting one quarter turn.

Remove the storage cap
a) Remove the pH probe storage cap by

gripping the top of the cap and gently

twisting the base one rotation clockwise to

loosen slightly. Next slowly slide the cap off

the pH probe.

DO NOT completely remove the base of the

cap from the top of the cap.

CAUTION: When the pH probe is not in use, add

enough Bluelab pH Probe KCl Storage Solution to

the storage cap so the probe tip is covered. Then

replace the cap and store in a secure place.

DO NOT use RO (Reverse Osmosis), Distilled

or De-ionized water. Pure water changes the

chemistry in the reference, causing the probe

to die.

Calibrate the pH
Calibrate the combo meter by following the

instructions in section 3.0 of this manual.
This must be done before the

Combo Meter is used for the fi rst time.

1

2

3

4

Attaching the

Bluelab pH Probe to the Meter

Removing pH probe

storage cap

Ensure probe tip is covered by

the KCl storage solution in cap

Inserting

Twisting

Attached

See section 3.0
for calibration
steps