Cleaning 11 battery replacement, 0 cleaning, 0 battery replacement – Bluelab Soil pH Pen User Manual
Page 11: English
English
11
9.0 Cleaning
To ensure accurate readings the probe tip needs to be kept clean at all times.
Cleaning prior to calibration is necessary for successful calibration.
Remove storage cap/dibber.
Rinse probe tip under fresh tap water.
Fill small plastic container with clean water.
Add a small amount of Bluelab pH Probe Cleaner or
mild detergent (dishwashing liquid).
Gently stir the probe tip in the mixture.
Ensure that you do not ‘knock’ the probe on the side of
the container as this may cause damage to the probe.
Rinse well under fresh running water to remove all
traces of the detergent mixture.
If the probe requires removal of heavy
contamination:
Gently brush around the glassware with a few drops
of Bluelab pH Probe Cleaner or mild detergent
(dishwashing liquid) and a soft toothbrush.
Rinse well under fresh running water to remove all
traces of the detergent mixture.
Calibration of the probe is required after every clean.
See the pH calibration in section 11.0. Place storage
cap/dibber back onto the probe tip.
10.0 Battery replacement
The pH pen is powered with 1 x AAA alkaline battery. Do not use rechargeable
batteries. A low battery warning is indicated by a battery symbol appearing on
the screen. Only remove the battery cap when the batteries require changing.
Battery life is expected to be approx. 350 hours.
To remove old battery
Undo battery cap fasteners. Remove battery cap and tip out the
old battery.
Fit new battery
Insert the new battery positive (+) end down into the body.
Ensure waterproof battery cap seal is clean, and free from
any debris.
Seal will fail if any dirt or debris is present.
Replace battery cap
Tighten fasteners on battery cap. Do not over tighten.
Ensure silicone seal on battery cap is fully enclosed in pen
body. This ensures the unit remains 100% waterproof.
Waterproof seal
1
2
3
4
5
6
1
2
3
4
probe cleaner
or dish
washing liquid
clean tap
water