9. Cleaning and storing the instrument
• Clean your device and cuff carefully only with a slightly mois-
tened cloth.
• The frequency of cleaning depends on the contamination
level of the device. Clean the blood pressure device and cuff
as soon as any dirt appears on the device.
• Do not use detergents or solvents.
• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
10. Rectifying faults
Error messages can occur when
• measurement error ( appears in the display),
• you move or talk during the measurement ( appears in the
display),
• the cuff tube is not properly inserted ( appears in the dis-
play),
• inflation takes longer than 15 seconds ( appears in the dis-
play),
• the inflation pressure is higher than 300 mmHg ( appears
in the display),
• an error occurs when storing the measured values ( ap-
pears in the display),
• exceeding measurement range ( appears in the display),
• the data could not be sent to the PC (
appears in the
display).
In the about cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do not
move or talk. Put the batteries back in if necessary, or else
replace them.
11. Specifications
Model no.
BM 75
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 30 – 260 mmHg,
diastolic 30 – 260 mmHg,
Pulse 40 –199 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 175 mm x W 117 mm x H 50 mm
Weight
Approx. 478 g (without batteries)
Cuff size
24 to 36 cm
Permissible operating
conditions
+10 °C to + 40 °C, 15 % – 85 % relative
air humidity (non-condensing)
Permissible storage
and transport condi-
tions
- 10 °C to + 60 °C, 10 % – 90 % relative
air humidity, 700 –1060 hPa ambient
pressure
Power supply
4 x 1,5 V
AAA batteries
Battery life
For approx. 120 measurements, de-
pending on the blood pressure level
and/or pump pressure