Monroe Electronics Resistivity Meter - Portable - Surface - model 272A User Manual
Page 6

6
RESISTANCE MEASUREMENTS:
Included with each Model 272A is a Model 96121-1 Surface-
to-Ground Measuring Adapter. This adapter used with the
probe and Model 272A meets requirements of the following
standards for measurement of resistance to ground of
flooring and work surface materials:
ASTM F-150
Standard Test Method for ELECTRICAL RESISTANCE OF
CONDUCTIVE RESILIENT FLOORING
ESD ASSOCIATION STANDARD NO. 4.1
Worksurfaces - Resistive characterization
ESD ASSOCIATION STANDARD NO. 7.1
Floor Materials - Resistive characterization of materials
NOTE: The adapter serves as a protective cap for the probe during storage and must be
removed to expose the concentric ring electrodes for resistivity measurements.
ASTM standards are available from:
American Society for Testing and Materials
1000 Barr Harbor Drive
West Conshohocken, PA 19428-2959
Phone (610) 832-9500
Fax (610) 832-9555
ESD Association standards are available from:
ESD Association, Inc.
7902 Turin Road
Suite 4
Rome, NY 13440-2069
Phone (315) 339-6937
Fax (315) 339-6793
Monroe Electronics, Inc. does not supply copies of standards.
To make a measurement of resistance from a surface to a ground point, connect the jack
labeled “OHMS” to the ground point, set the electrode on the surface and switch to the “OHMS”
mode (see Figure 2).
Surface-to-surface measurements, also directed in these standards, require the use of a second
electrode (Monroe Electronics Model 96083) as shown in Figure 2. Model 96083 electrodes are
included in Models 96085-1 and 96085-2 electrode sets or they may be ordered separately.
Figure 2
Setup for Surface-to-Ground and Surface-
to-Surface
Resistance Measurements
272A
SET TO INDICATE
96101A
96121-1
OR
WORKSURFACE
96083