Microphone type
Coupler required
Correction to nominal level dB
CEL-186/2P CEL-2835
-1.0
CEL-186/2PR CEL-2835
-1.0
CEL-192F CEL-2835
-1.0
CEL-190F Not
required
-1.0
Genrad 1961
Not required
0
Genrad 1962
CEL-2834
0
Knowles BL1803
CEL-2834
0
B&K 4133
CEL-2835
-1.0
Figure 1
Microphone equivalent volume corrections to nominal output sound level
CEL-177 Acoustic Calibrator
level variation with altitude
-6.00
-5.00
-4.00
-3.00
-2.00
-1.00
0.00
1.00
2.00
-5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9 1011121314151617181920
Altitude (ft x 1000)
Correction to output
level (dB)
Figure 2
Calibration corrections as a function of altitude realtive to sea level
CEL Dosimeter
model number
Dosimeter calibration count at
nominal output 114 dB
(includes coupler where used)
CEL-139D
58.2
CEL-172H & J
58.2
CEL-172K
184.8
CEL-172L
116.4
CEL-179H & J
46.2
CEL-179K
146.7
CEL-179L
92.4
CEL-180G
46.2
CEL-180J
146.7
CEL-180L
5.5
CEL-181G
58.2
CEL-181J
184.8
CEL-181L
6.3
Figure 3
Dosimeter calibration settings using CEL-177 Acoustic Calibrator