Casella CEL Thermohygrograph standard model User Manual

Page 5

background image

3.

PREPARATION FOR USE

Once you have removed the THERMOHYGROGRAPH standard from the

packing you should have

¤

THERMOHYGROGRAPH standard,

¤

Tool kit,

¤

Document showing settings for the clock,

¤

Chart drum with lid,

¤

User Manual,

¤

Calibration certificate (supplied only with M109015).

Make sure that all foam packing is removed from inside the

THERMOHYGROGRAPH case. Also make sure that the rubber band which

holds the pen arms together is removed.

3.1

Removing/Replacing the Drum

Before attempting to remove or replace the drum, it is important to move the

pens out of the way by pulling the pen lifter control outward.

To remove the chart drum, first open the top of the unit. Then take off the lid

of the drum by simply pulling it upwards. The drum can now be pulled out by

holding the inside walls of the drum and lifting. Note that the drum may be

fairly tight in a new unit.

To replace the drum, carefully locate it on the clock drive shaft and firmly

push it home.

3.2

Fitting/Replacing the Battery

After removing the drum, the battery access cover can be found on the side

of the clock, at the rear of the unit.

To open the battery compartment, use either a coin or flat headed

screwdriver to rotate the cover a quarter turn anti-clockwise. The clock uses

a single AA (MN1500 or LR6) alkaline battery, which should be inserted with

the positive end facing outwards. To refit the battery cover, push it into the

battery compartment, ensuring that the slot is vertical. Turn the cover a

quarter turn clockwise to lock it.

3.3

Setting the Clock speed

With the chart drum removed set the three dip-switches in the clock base to

suit your requirements.

Preparation

Page 5 of 12

THERMOHYGROGRAPH standard
User Manual