Casella CEL CEL-278 User Manual
Page 28

Typical noise floor: (measured on Lo range with a CEL-493/2 Sound Level
Meter)
Filter Set All pass: (WTNG and band position)
Linear
A-weighted
28.0dB 15.5dB(A).
Filter Set C-weighted:
Linear
21.0dB.
Octave bands:
Band
Linear
A-weighted
16Hz
19.0dB <
10.2dB(A)
31.5Hz 18.0dB <
10.2dB(A)
63Hz
14.0dB <
10.2dB(A)
125Hz
11.0dB <
10.2dB(A)
250Hz
10.5dB <
10.2dB(A)
500Hz
10.0dB <
10.2dB(A)
1kHz
10.0dB <
10.2dB(A)
2kHz
10.0dB <
10.2dB(A)
4kHz
10.0dB <
10.2dB(A)
8kHz
10.0dB <
10.2dB(A)
16kHz
11.0dB <
10.2dB(A)
Max Input voltage:
Lo range
Conditioned 0.25V RMS
Hi range
Conditioned 8.0V RMS.
Input impedance: 10kΩ,
Recommended output impedance of input device < 1 kΩ
Page 28
CEL-278/2 Handbook
CEL Instruments Ltd