Handbook No. HB3182-02 Issue2/JULY 1994
Possible cause:-
Service Hint.
Grub screw securing eccentric to motor
Re tighten
shaft loose.
Symptom:- Pump runs but no air drawn.
Possible cause:-
Service Hint.
Diaphragm rubber split.
Replace.
Pulsation damper membrane split.
Renew pulsation
damper.
Flow potentiometer on bottom setting.
Re-adjust.
Tube between pulsation damper and
Replace.
valve chest split or disconnected.
See also other documents in the category Casella CEL Measuring instruments:
- CEL-62X (37 pages)
- CEL-63X (87 pages)
- CEL-430 (8 pages)
- CEL-440 (70 pages)
- CEL-480 (2 pages)
- CEL-231 (2 pages)
- CEL-283 (6 pages)
- CEL-228 (6 pages)
- CEL-275 (56 pages)
- CEL-393 (99 pages)
- CEL-193 (20 pages)
- CEL-224 (4 pages)
- Vortex personal sampling pumps (14 pages)
- Vortex Ultraflow area sampling pump (12 pages)
- Tuff personal sampling pumps (38 pages)
- Tuff personal sampling pump (1 page)
- Environmental enclosure for Microdust (10 pages)
- Sampling heads for personal dust sampling (24 pages)
- Apex personal sampling pumps (40 pages)
- Apex personal sampling pump (2 pages)
- CEL-630 (2 pages)
- Airborne bacteria sampler (16 pages)
- CEL-620 (2 pages)
- CEL-500 (16 pages)
- CEL-400 (2 pages)
- CEL-450 (77 pages)
- CEL-272 (8 pages)
- CEL-420 (16 pages)
- CEL-320 (24 pages)
- CEL-300 (2 pages)
- CEL-268 (99 pages)
- CEL-350 (2 pages)
- CEL-35X dBadge (48 pages)
- CEL-315 (45 pages)
- CEL-180 (14 pages)
- CEL-281 (147 pages)
- Thermohygrograph standard model (12 pages)
- Tipping bucket rain gauge (16 pages)
- Thermohygrograph deluxe model (12 pages)
- Piche evaporimeter (4 pages)
- Nomad portable weather system (2 pages)
- Fortin barometer (11 pages)
- Campbell Stokes Sunshine recorder (2 pages)
- Barographs small and display pattern (2 pages)