American Weigh Scales (AWS) LS-110 User Manual
Page 6

LS-110 Manual
eral times until the inner code displays (~2680)
3. Press again. The display will flash “CAL” followed
by the required calibration weight (20kg).
4. Add the required weight. The display will show pass
and return to normal weighing mode.
5. Remove the weigh and press and hold the key to
turn the scale off. Calibration is complete!
IX. Troubleshooting
•
Problem: Scale will not turn on
•
Solution: Check the batteries and ON/OFF lock
•
Problem: Display will not stabilize
•
Solution: Make sure the weight is stable on the
hook and not shaking or swinging
•
Problem: Displayed weight is inaccurate
•
Solution: Perform a calibration adjustment
X. Error Codes
1. OUTZ - Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. EEEEE - Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. LLLLL - Load cell has been overloaded.
4. UNST - The scale is unstable. Use the scale in a
more stable position