American Weigh Scales (AWS) LB-1000 User Manual
Page 6
![background image](https://www.manualsdir.com/files/766277/content/doc006.png)
LB-Series Manual
5. Press
TARE/PRINT again to zero out the balance
and the scale is ready for normal weighing again.
IX. Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
LB-1000
0.2g
400g
OIML M2
LB-3000
0.2g
1000g
OIML M2
X. Adjustment
1. Press and hold the
MODE/CAL key until the display
shows “CAL”.
2. Press the
MODE/CAL key again. The display will
flash “CAL” followed by the calibration weight value
(see chart).
3. Place the calibration weight on the center of the
weighing platform. Make sure you are using the
correct weight for your specific model.
4. Once the weight is accepted, the display will show
“PASS” then return to normal operation. Calibration
is complete!