VI. Auto-Off Feature

This scale has an Auto-Off feature that can help extend battery

life by turning the scale off when not in use for a set amount

of time (1-9 minutes). To change the Auto-Off timer, press the

“Menu” button three times. The scale will show the current

Auto-Off setting (OFF, 60, 120, 180, 240, 300, 360, 420, 480,

or 540)

Press the “Unit” button to cycle through until you find the

desired timer duration (in seconds). Press “Tare” to confirm

your selection.

VII. Calibration

This scale is calibrated at the factory, but frequent use or

slight shock can sometimes affect the calibration. In such

cases you may need to recalibrate the scale. Typically, the

Standard Calibration method is sufficient for the average

user to recalibrate their scale. Advanced calibration is already

performed at the factory and should not be attempted unless

you have two calibration weights. Before starting calibration,

insure that the area you are working on is free of any drafts,

vibrations, and you are using a flat level surface

• Standard Calibration (Span):

**MiniPRO-50 calibrates with a 20g or 50g weight.

**MiniPRO-100 calibrates with a 50g or 100g weight.

Press the “Menu” button one time, “SCALE” will be displayed

on the LCD; Press the “Tare” button to continue. The scale will

display the required calibration weight. You can press “Unit”

to change between the two available calibration weights (see

above). Press “Tare” to confirm. The value on the display will

flash. Place your calibration weight onto the weighing platform.