American Weigh Scales (AWS) SM-500 User Manual

Page 5

background image

4. To return to normal weight display (gross weight),

remove any items from the weighing platform and

press the

TARE key again to re-zero the scale.

VII. Calibration
Calibration and adjustment are performed at the factory.

However, the weighing range can shift slightly during

shipping. Also, due to changes in local acceleration of gravity

around the world, adjustment is recommended when moving

the scale from one location to another. To check calibration:

1. Place the required calibration weight on the scale

and note the displayed value.

2. If the displayed value is not within acceptable

tolerance (see chart below), adjustment is required.

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

SM-200

0.02g

200g*

OIML M2

SM-500

0.2g

500g*

OIML M2

SM-501

0.02g

500g*

OIML M2

SM-DR

0.2g

500g*

OIML M2

*Can be changed to 100g during calibration by pressing

TARE. See Section VIII.

VIII. Adjustment

1. Turn the scale on by pressing the ON key.
2. Wait for the display to show “0.0g” or “0.00g” then

press and hold the ON key for a few seconds until

the display shows “0”, then release.

3. The display will flash “g” along with the default

calibration weight value (see chart above). Place the