4. To return to normal weight display (gross weight),

remove any items from the weighing platform and

press the

T key again to re-zero the scale.

VII. Calibration
Calibration and adjustment are performed at the factory.

However, the weighing range can shift slightly during

shipping. Also, due to changes in local acceleration of gravity

around the world, adjustment is recommended when moving

the scale from one location to another. To check calibration:

1. Place the required calibration weight on the scale

and note the displayed value.

2. If the displayed value is not within acceptable

tolerance (see chart below), adjustment is required.

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

BT2-201

0.01g

200g, 100g*

OIML M2

BT2-1KG

0.1g

500g, 100g*

OIML M2

*Selectable at calibration time

VIII. Adjustment

1. Press and hold the

M key until the display shows

“CAL”.

2. Press the

M key again. The display will flash “CAL”

followed by the calibration weight value (see chart).

3. Place the calibration weight on the center of the

weighing platform.

4. Once the weight is accepted, the display will show

“PASS” then return to normal operation. NOTE: If