1. Place the empty container or vessel on the weighing

platform.

2. Press the

TARE key once. The display will then

reset to zero. If the container is removed, it’s weight

will be displayed as a negative number.

3. You may now fill the container with the quantity to

be weighed. The display will show the net weight of

the contents.

4. To return to normal weight display (gross weight),

remove any items from the weighing platform and

press the

TARE key again to re-zero the scale.

VII. Calibration
Calibration and adjustment are performed at the factory.

However, the weighing range can shift slightly during

shipping. Also, due to changes in local acceleration of gravity

around the world, adjustment is recommended when moving

the scale from one location to another. To check calibration:

1. Place the required calibration weight on the scale

and note the displayed value.

2. If the displayed value is not within acceptable

tolerance (see chart below), adjustment is required.

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

AMW-70

0.02g

50g

OIML M2

AMW-100

0.02g

100g

OIML M2

AMW-250

0.2g

200g

OIML M2

AMW-600

0.2g

500g

OIML M2

AMW-1KG

0.2g

500g

OIML M2