Installation – ABUS Technologies myPCLab Data Acquisition / Recorder User Manual
Page 8

ABUS TECHNOLOGIES INC.
8
myPClab
3.7 Voltage Digital Input Wiring
Digital signals are connected to channel 3, terminals 1 and 2. Observe
indicated polarity. Voltage levels applied to this channel must be within the
specified limits. See “Technical Parameters”.
5. INSTALLATION
5.1 Recommendation
myPCLab inputs are not electrically isolated from the USB port. Before connecting
signals from sensors or transducers, make sure that they are not in contact with any source of
electrical power (wires or conductive surfaces which may get energized). Input channels are not
isolated from each other.
5.2 Input Signal
INPUT TYPE
MAXIMUM RANGE
J Thermocouple
-130,0 to 940,0 °C (-202,0 to 1724,0 ºF)
K Thermocouple
-200,0 to 1370,0 °C (-328,0 to 2498,0 ºF)
T Thermocouple
-200,0 to 400,0 °C (-328,0 to 752,0 ºF)
E Thermocouple
-100,0 to 720,0 °C (-148,0 to 1328,0 ºF)
N Thermocouple
-200,0 to 1300,0 °C (-328,0 to 2372,0 ºF)
R Thermocouple
0,0 to 1760,0 °C (-32,0 to 3200,0 ºF)
S Thermocouple
0,0 to 1760,0 °C (-32,0 to 3200,0 ºF)
B Thermocouple
500,0 to 1800,0 °C (932,0 to 3272,0 ºF)
Pt100 RTD
-200,0 to 650,0 °C (-328,0 to 1202,0 ºF)
0 to 50 mVdc
Adjustable from -32767 to +32767
0 to 10 Vdc (Channel 1)
Adjustable from -32767 to +32767
4 to 20 mAdc
Adjustable from -32767 to +32767
myPCLab module sensors and signals