Fluke 87 V User Manual
Page 36

80 Series V
Users Manual
28
Table 6. Functions and Trigger Levels for Frequency Measurements
Function Range
Approximate
Trigger Level
Typical Application
K
6 V, 60 V,
600 V, 1000 V
± 5 % of scale
Most signals.
K
600 mV
± 30 mV
High-frequency 5 V logic signals. (The dc-coupling of the L function can
attenuate high-frequency logic signals, reducing their amplitude enough to
interfere with triggering.)
mL
600 mV
40 mV
Refer to the measurement tips given before this table.
L
6 V
1.7 V
5 V logic signals (TTL).
L
60 V
4 V
Automotive switching signals.
L
600 V
40 V
Refer to the measurement tips given before this table.
L
1000 V
100 V
R
e E Gd
Frequency counter characteristics are not available or specified for these functions.
\
All ranges
± 5 % of scale
AC current signals.
μ
AF
600
μ
A, 6000
μ
A 30
μ
A , 300
μ
A
Refer to the measurement tips given before this table.
^
60 mA, 400 mA
3.0 mA , 30 mA
AF
6 A, 10 A
0.30 A, 3.0 A