Appendix B: CALIBRATION & SERVICE
It is recommended that the instrument be calibrated every 12 months. Please
consult Test Products International for further details.
The following are consumable parts for the instrument:
In-Line Filter Element (pkg of 5)
User Replaceable
A794F
Mini Pump Protection Filter Assem. User Replaceable
A763
**Oxygen Sensor
User / Factory Replaceable
A761
**Carbon Monoxide Sensor
User / Factory Replaceable
A760
**Sensor replacement requires calibration gas.
The following are accessories for the instrument:
Flue Temperature Probe w/k-type
Standard
A770
'K' Type Temperature Probe
Standard
GK11M
In-Line Filter Complete
Standard
A794
Mini Pump Protection Filter Assem. Standard
A763
Rubber Boot
Standard
A765
Infrared Printer
Optional
A740
Soft Carrying Case
Standard
A768
Exhaust Spigot
Standard
A764
Appendix C: GUARANTEE
Your TPI 709 Flue Gas Analyzer is guaranteed free from defects in materials
and workmanship for 3 Years from the date of purchase. This guarantee does
not affect your statuary rights. For additional information please refer to the
included warranty card or contact TPI at 800-368-5719.
To obtain warranty performance or maintenance on your analyzer: - Include
with the product your name, address, phone number, written description of
the problem and proof of purchase date. Carefully package and return to:
TPI / Attn. Repair
9615 SW Allen Blvd. Suite 104
Beaverton, OR 97005
Appendix A : SPECIFICATIONS (Continued)
Pressure Measurement
Selectable Ranges
mbar, kPa and inH2O
Range
- 150 mbar to + 150 mbar
-15 kPa to + 15 kPa
-60 inH
2
O to 60 inH
2
O
Resolution
0.01 mbar, 0.001 kPa, 0.001 inH
2
O
Accuracy
+/- 0.5% fsd
Temperature Measurement
Input Type
K-Type thermocouple
Range
-58°F to 1832°F (-50°C to 1000°C)*
Resolution
1°F (1°C)
Accuracy
+/- (0.3% of rdg + 2°F) or +/- (0.3% of rdg +
1°C)
* The thermocouple supplied (GK11M) has the ability to measure tempera-
tures in the -50°F to 950°F range. The 712 accepts industry standard K-type
thermocouple probes with sub-mini connectors.