beautypg.com

REFCO REF-VAC User Manual

Page 4

background image

2

EN

Technical details

Property

Values

Range

0 to 18000 Microns / 0 to 2400 Pascal

Units

Pascal, Microns, mbar, Torr, mTorr, psi, inHg

Resolution

0 – 20:

0.2 Microns

21 – 200:

1 Microns

201 – 500:

5 Microns

501 – 1000:

10 Microns

1001 – 3000:

50 Microns

3001 – 6000:

200 Microns

6001 – 18000:

1000 Microns

Sensor type

Thermistor

Connector

1/4 SAE = 7/16-20UNF

Operating

temperature range

0°C to 40°C / +32°F to 104°F

Protection class

IP54

Storage temperature

range

-20°C tp +60°C / -4°F to 140°F

Max. overpressure

34 bar / 500 psi

Accuracy

up to 100 Microns +/- 10 Microns
101 - 750 Microns +/- 45 Microns

Power supply

3 x 1.5 V AAA / LR03 batteries
Service life when used continually is ca. 40 hours

(depending on the battery type).

Automatic shut off

After 10 min. if vacuum value greater than 18000

Microns
After 30 min. if vacuum value less than 18000 Microns

Use

Preparation

Insert 3 batteries in the battery compartment at the back of the device.

Ensure the batteries are inserted using the correct polarities.

Do not leave used batteries in the battery compartment. If you

will not be using the device for a longer period of time, remove

the batteries from the battery compartment.