Appendix 1: ADJUSTING CONTROL PARAMETERS
(PAK REACH 400 METER IN TCAT-2 CONTROLLERS)
7
QX EB TCAT-2DF 06/24/05 PAK 400
GENERAL SETUP PROCEDURE
Turn the system power ON
Do not connect the load at the controlled outuput at this time.
At first, as the power turns ON, the meter will display in the top part “PV”....”in IE” and in the lower part
“SV”....0.0.. This is an indication that the feedback thermocouple is not connected to the system.
Press SET button for several seconds until top”PV display shows “P! and bottom “sv” shows some num-
bers.
Continue pressing SET several time until “PV” displays show LCk and ”sv” shows some numbers. Change
these ‘sv” numbers to 1111.
Press SET and SHIFT buttons together for several seconds to set ten function setpoints.
STEP 1
The word “SET1” appears in PV and sv displays some numbers. Reset the numbers in the sv column as
listed below correspondeing to the parameter shown in Pv. Continue resetting set points SEt1...to SEt0.