Physitemp Instruments BFS-30MP User Manual
Page 5
6.0
SPECIFICATIONS
6.1
CONTROLLER
Operating Range:
BFS-3MP and 5MP:
-35°C to ambient
BFS-30MP:
-40°C to ambient
Digital Readout Resolution:
0.1°C
Accuracy:
±0.2°C
Ambient Operating Range:
15° - 45°C
Input Power Requirements:
100 - 120VAC @ 4 Amps or
200 - 240VAC @ 2 Amps
Grounded Outlet
Power Output to stage:
BFS-30MP - 12VDC @ 10Amps
BFS-3 & 5 MP - 5VDC @ 10Amps
Display:
Dual LED display of set and run Temperature
Safety Features:
Power shutoff during over-temperature condition caused by
inadequate coolant flow (50°C). Short circuit protection.
Size:
9 1/2” wide x 10” deep x 4 1/2” high
Weight:
5 lbs
Other Features:
Self-sealing water connectors.
Auxiliary switched AC output for ECV-1 or PTU-3.
6.2
STAGE
Temperature Range:
-40°C to ambient
Controlled Surface Area:
BFS-3MP
- 3 x 4cm
BFS-5MP
- 3.8 x 4cm
BFS-30MP
- 7.5 x 8cm
Weight:
15ozs, including leads
Lead Length:
5ft
8
EB BFS-MP 6310 04/04