Physitemp Instruments BAT-10 User Manual
Page 6

10.0
SPECIFICATIONS
Temperature Range
-200°C to +400°C, 1°C resolution
and Resolution:
-100°C to +199.9°C, 0.1°C resolution
Differential
-19.99°C to +19.99°C. Linearization centered at 40°C
Temperature Range:
See Section 5.0
Accuracy: 1° Range
1°C, (±1 digit)
0.1° Range
0.1°C, (±1 digit)
Diff. Range
0.01°C (±1 digit)
Repeatability:
One least significant digit
Calibration:
Conforms to National Institute of Standards and Technology
(Monograph 125)
Sensors:
Any Physitemp type T Thermocouple
Ambient Operating Range:
15 - 45°C
Readout:
3 1/2 digits
Batteries:
BAT-10 - 4 alkaline “C” cells (Battery life, 1000 hours)
BAT-10R - (rechargeable unit) - 4 NI-Cads
Special Display
LO-BAT appears on display when battery voltage is low, but
Indications:
several hours of life remain.
INPUT appears on display when open probe or input
fault condition has occured.
ALARM, acitivated by user supplied external 5V logic level
Analog output:
Non-linearized set at 1.5V corresponding to temperature of 401°C
Linearized (when supplied) 1mV/count
7