Physitemp Instruments BAT-7001 User Manual
General introduction, Operating instructions
GENERAL INTRODUCTION
MODEL BAT-7001H is a high performance precision hand-
held instrument for the measurement of temperature using
type T (Copper-Constantan) Thermocouples. This versatile
instrument measures temperature in both °C and °F. The
internal memory of the micro-computer stores the minimum
and maximum measured temperature for instant recall via
the LCD display. BAT-7001H operates on a 9V battery.
INITIAL INSPECTION
Unpack carefully and inspect for any damage from shipment.
If damage is found, notify Physitemp immediately. All packing
materials should be saved until satisfactory operation is con-
firmed.
SPECIFICATIONS
Range:
-90 to 400°C
-130 to 752°F
Resolution:
0.1°C
0.2°F
Accuracy:
±0.05% of reading. ±0.7°C or 1.3°F
Accuracy is specified for operating ambient
temperature of 23°C ±5°C
Cold junction error for thermocouples: For
operating ambient temperature from 0°C to
+18° and +28°C to 50°C, add an additional
error of ±0.03°C/°C or 0.03°F/°F
OPERATING INSTRUCTIONS
TOUCH KEYS:-Refer to Figure 1, above
1.
The C/F key selects the readout for °C or °F
2.
The readout will alternately display the maximum
and minimum values when MIN/MAX is pressed.
3.
The RESET key clears the minimum and maximum
values stored in the memory of the micro-computer.
ERROR CODE:- Refer to Figure 1, above
1.
HELP : Open sensor indication
2.
HELP
: Over range indication
3.
-HELP : Under range indication
POWER SWITCH
PROBE
LCD
TOUCH KEYS
ADAPTOR INPUT