Mtm installation, Module – Parr Instrument MTM User Manual
Page 6

MTM Installation
P a r r I n s t r u m e n t C o m p a n y
6
Module
Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select
type
Main Screen: SV = 100
Press return key and release
Operation
Mode
Select
type/value Comment
r-S
Run
Run/Stop
SP
1
Decimal point position
LoC
OFF
Lock mode (lock all
keys or only up/down
arrow able to use)
Press and hold down “SET” for 5-sec
Operation
Mode
Select
type/value Comment
InPt
v10
Input type (v5 = 0-5V,
v10 = 0-10V, nA0 =
0-20mA, nA4 = 4-20mA,
nV =0-50mV)
tP-H
To be de-
termined*
Upper-limit range
tP-L
0
Lower-limit temperature
range
CTRL
ON/OFF
Control mode (ON/OFF,
MANUAL, PID and PID
PROG)
S-HC
Heat
Heat/Cool control
ALA1
0
Alarm mode
SALA
OFF
System alarm
CoSH
ON
Communication write
function that able to use
set point from software
C-Sl
ASCII
Format type
C-no
See note** Communication address
bPS
9600
Communication baud
rate
Len
7
Data length setting
Operation
Mode
Select
type/value Comment
PrtY
Even
Parity bit setting
StoP
1
Stop bit setting
* For 90VDC motor, 115VAC, tP-H = 142.9
For 180VDC motor, 230VAC, tP-H = 120.4
For 90VDC motor, 230VAC, tP-H = 153.2
** First available communication port (3 or 4)
Press “SET” and release
Operation
Mode
Select
type/value
Comment
HtS
0
Hysteresis
TPoF
0
Inaccuracy adjustment
CrHI
0
Analog output low
CrLo
0
Analog output high