High temperature module - htm – Parr Instrument HTM User Manual
Page 6

- 6 -
HTM Installation
High Temperature Module - HTM
Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select type
Main Screen: SV = 375
Press return key and release
Operation
Mode
Select
type/value
Comment
r-S Run Run/Stop
SP
0**
Decimal point position
AL1H 0
Upper-limit
alarm
1
LoC
OFF
Lock mode(lock all keys or only up/down arrow able to use)
**
If using a type-J or type-K thermocouple, SV = 0
If using an RTD, SV = 1
Press and hold down “SET” for 5-sec
Operation
Mode
Select
type/value
Comment
InPt J** Input
type
tPUn
C
Temperature unit
tP-H
800
Upper-limit temperature range
tP-L
0
Lower-limit temperature range
CTRL
ON/OFF
Control mode (ON/OFF, MANUAL, PID and PID
PROG)
S-HC Heat Heat/Cool
control
ALA1
2
Alarm operates when PV value is higher than SV value +
ALH setting
SALA OFF
System
alarm
CoSH
ON
Communication write function that able to use set point from
software
C-Sl ASCII Format
type
C-no 2
Communication
address
bPS
9600
Communication baud rate
LEn
7
Data length setting
PrtY
Even
Parity bit setting
StoP 1
Stop
bit
setting
**
If using a type-J thermocouple, InPt = J
If using a type-K thermocouple, InPt = K
If using an RTD, InPt = Pt