beautypg.com

Pdm installation – Parr Instrument PDM User Manual

Page 4

background image

PDM Installation

P a r r I n s t r u m e n t C o m p a n y

4

Wiring Installation in 4848 Reactor
Controller (Continued)

9. Find the free orange wire #1 from the kit and at-
tach one end to pin 8 on the 2082E meter. Take care
not to apply too much torque which could cause the
wire to break. (This may already be connected for
your convenience.) The other end attaches to termi-
nal block position #6.

10. Find the free blue wire #2 from the kit and attach
one end to pin 7 on the 2082E meter. Take care not
to apply too much torque which could cause the
wire to break. (This may already be connected for
your convenience.) The other end attaches to termi-
nal block position #7.

Pin Outs:

2082E

Color:

Attaches to:

Pin 1

Black

Terminal Block 5

Pin 2

White

Terminal Block 2

Pin 3
Pin 4

White

A2104E Press Harness

Pin 5
Pin 6

Black

A2104E Press Harness

Pin 7

Blue

Terminal Block 7

Pin 8

Orange

Terminal Block 6

Pin 9
Pin 10
Pin 11

White

Terminal Block 4

Pin 12

Black

Terminal Block 3

Final Steps:

Close the controller and replace the two screws on
the top plate. Plug the 4848 controller back in, and
turn it on.

It is useful to check that the settings on the display
are set correctly. Check these against the defaults
listed in the back of these instructions.