5 Calibration - Using an External Calibration
Weight
Calibration is required when the weighing machine is initially installed or if the
Balance is moved to a substantial distance from the original location. 30
minutes of warming up of the Balance is always needed before weighing and
calibration.
Turn the Balance on and when display shows
“--------”, press ZERO
key
The display will show “CAL”
Press ZERO key again, the display w
ill show “X000.00”, “X” is the
flashing digit
Press UNIT key, the flashing digit will move to right
Press TARE key to increase the value of the flashing digit(“X” is the
calibration weight can be set according to users’ requirements, we
recommend a minimum weight of at least 50% of the Balance capacity)
Press ZERO key, display will show “CAL” and then series of the digit,
wait for stable indication shows, press UNIT key , display will show
--------
, then “X000.00”
Place the corresponding test weight on the center of the pan
The display backlight activated and returns to weighing mode
Now the calibration is completed
Turn off the power, then turn on the power again, place a known weight
on the platform to verify if it is accurate.
If not, repeat above steps.