3 getting started with your weighing machine, 1 unpacking the equipment – LW Measurements LWC 800 User Manual
Page 14

14
Transmission Information Format: 20 Byte ,blank=20H
1~2
3
4~13
14~18
19
20
‘W: ’
P
DATA
UNIT
CR
LF
 
W: 
Start of Data Transmission=57H+3AH
P :
Polarity
‘+’ = 2BH = Positive
‘-’ = 2DH =Negative
DATA: 12.345’’=20H+20H+20H+20H+31H+32H+2EH+33H+34H+35H 
UNIT: 
‘lb’
=6CH+62H+20H+20H+20H
‘kg’
=6BH+67H+20H+20H+20H
CR:
= 0DH
LF:
= 0AH
 
If you have any questions on the technical data or require detailed technical 
information on your balance, please contact your technical representative. 
 
3 Getting started with your weighing machine
3.1 Unpacking the equipment
The machine is delivered in an environmentally-friendly carton, specifically 
developed for this precision instrument, which provides optimum protection for 
the machine during transportation. 
 
We suggest that you retain the original packaging in order to avoid 
transportation damage if re-shipping or transporting the machine and to allow 
the unit to be stored in the best conditions if it is out of operation for an 
extended period. 
 
In order to avoid damage, attention must be given to the following points when 
unpacking the weighing machine: 
 
Unpack the weighing machine carefully. It is a precision instrument.
When outside temperatures are very low, the machine should first be 
stored for some hours in the unopened transport package in a dry 
room at normal temperature, so that no condensation settles on the 
unit when unpacking.
