4.2 Program options
usage.
4.2.1 Interface Settings
Press [ON/OFF] to turn on the scale, when instrument displays --------
Press [PRINT], the display will show b xxx BAU value
Press [UNIT] to select among (1200,2400,4800,9600), default setting is 9600
Press [ZERO] to enter Parity mode, the display will show P xxx PAR
Press [UNIT] to select odd, even or none ( odd stands for 7 data bits with odd parity, even stands
for 7 data bits with even parity, none stands for 8 data bit without parity), default setting is none
Press [ZERO] to enter Print mode, the display will show n xxx mod
Press [UNIT] to select CON,OFF, KEY, STB
a)
OFF stands for serial data output disabled
b)
KEY stands for manual mode by pressing [PRINT]
c)
STB stands for automatic print when scale is stable
d)
CON stands for continuous print
Default setting is KEY
Press [ZERO] again to return to weighing mode.
4.2.2 Parameter settings
4.2.2.1 Setting zero tracking range
Press ON/OFF key, and when
“- - - - - -
“ shows, press PCS key, display will show “Zrd x.x”, press UNIT
key to select, factory default setting is 0.8
4.2.2.2 Setting zero tracking time
Press ON/OFF key , and when
“- - - - - -
“ shows, press PCS key, and then press ZERO key till display
shows
“Zrt x.x”, press UNIT key to select, factory default setting is 2.0