Model 38300
Sash Height (inches)
14
Downflow (FPM)
50-55
Approximate Settings for new Filters
Mag Gauge Reading
0.3-0.4 upper / 0.5-0.7 lower
Supply Air Volume Displacement (CFM)
440
Number of Laskin Nozzles Needed
1
Exhaust Air Volume Displacement (CFM)
650-670
Number of Laskin Nozzles Needed
2
Supply HEPA Filter Dims. (in.)
48x18x3
Exhaust HEPA Filter Dims (in.)
48x18x6
Fluorescent Light(s)
1 each F32T8/TL741
*For further details on the secondary inflow method,
consult the inflow velocity calculation – Secondary
Method section of Chapter 6: Maintaining Your
PuriCare.
Use the downflow velocity test grid as described in this
manual when checking the downflow air velocity.